Standard Practice for Alternate Actinide Calibration for Inductively Coupled Plasma-Mass Spectrometry

ASTM International, 01/01/2014

Publisher: ASTM

File Format: PDF

$31.00$62.00


Published:01/01/2014

Pages:6

File Size:1 file , 98 KB

Note:This product is unavailable in Russia, Ukraine, Belarus

1.1 This practice provides guidance for an alternate linear calibration for the determination of selected actinide isotopes in appropriately prepared aqueous solutions by Inductively Coupled Plasma-Mass Spectrometry (ICP-MS). This alternate calibration is mass bias adjusted using thorium-232 (232Th) and uranium-238 (238U) standards. One of the benefits of this standard practice is the ability to calibrate for the analysis of highly radioactive actinides using calibration standards at much lower specific activities. Environmental laboratories may find this standard practice useful if facilities are not available to handle the highly radioactive standards of the individual actinides of interest.

1.2 The instrument response for a series of determinations of known concentration of 232Th and 238U defines the mass versus response relationship. For each standard concentration, the slope of the line defined by 232Th and  238U is used to derive linear calibration curves for each mass of interest using interference equations. The mass bias corrected calibration curves, although generated from interference equations, are specific to the instrument operating parameters and tuning in effect at the time of data acquisition. Because interference equations are part of the normal ICP-MS manufacturer's software package, this calibration methodology is widely applicable.

1.3 For this standard practice, the actinide atomic mass range that has been studied is from amu 232-244. Guidance for an extended range of amu 228-248 is given in this practice.

1.4 Using this practice, analyte concentrations are reported at each amu and not by element total (that is, 239Pu versus plutonium).

1.5 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

More ASTM standard pdf

ASTM D6426-04

ASTM D6426-04

Standard Test Method for Determining Filterability of Middle Distillate Fuel Oils

$30.00 $60.00

ASTM F773M-16

ASTM F773M-16

Standard Practice for Measuring Dose Rate Response of Linear Integrated Circuits (Metric)

$24.00 $48.00

ASTM B232M-97

ASTM B232M-97

Standard Specification for Concentric-Lay-Stranded Aluminum Conductors, Coated Steel-Reinforced (ACSR) [Metric] (Withdrawn 1999)

$30.00 $60.00

ASTM D5147-02ae1

ASTM D5147-02ae1

Standard Test Methods for Sampling and Testing Modified Bituminous Sheet Material

$30.00 $60.00