Your shopping cart is empty!
ASTM International, 06/01/2015
Publisher: ASTM
File Format: PDF
$30.00$60.00
Published:01/06/2015
Pages:8
File Size:1 file , 160 KB
Note:This product is unavailable in Russia, Ukraine, Belarus
1.1 This test method covers a gas chromatographic procedure for the quantitative determination of organic oxygenated compounds in gasoline having a final boiling point not greater than 220 °C and oxygenates having a boiling point limit of 130 °C. It is applicable when oxygenates are present in the 0.1 % to 20 % by mass range.
1.2 This test method is intended to determine the mass concentration of each oxygenate compound present in a gasoline. This requires knowledge of the identity of each oxygenate being determined (for calibration purposes). However, the oxygen-selective detector used in this test method exhibits a response that is proportional to the mass of oxygen. It is, therefore, possible to determine the mass concentration of oxygen contributed by any oxygenate compound in the sample, whether or not it is identified. Total oxygen content in a gasoline may be determined from the summation of the accurately determined individual oxygenated compounds. The summed area of other, uncalibrated or unknown oxygenated compounds present, may be converted to a mass concentration of oxygen and summed with the oxygen concentration of the known oxygenated compounds.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Methods for Chemical Analysis of Chromium and Ferrochromium
$27.00 $54.00
Standard Specification for Low-Alloy Steel Deformed and Plain Bars for Concrete Reinforcement
$30.00 $60.00
Standard Specification for Metallic Mechanical Fittings for Use on Outside Diameter Controlled Thermoplastic Gas Distribution Pipe and Tubing
Methods for Detection and Enumeration of Microbiological Contaminants in Water Used for Processing Electron and Microelectronic Devices
$106.00 $213.01