• ASTM D7315-17

ASTM D7315-17

Standard Test Method for Determination of Turbidity Above 1 Turbidity Unit (TU) in Static Mode

ASTM International, 07/01/2017

Publisher: ASTM

File Format: PDF

$37.00$75.00


Published:01/07/2017

Pages:22

File Size:1 file , 700 KB

Note:This product is unavailable in Russia, Ukraine, Belarus

1.1 This test method covers the static determination of turbidity in water. Static refers to a sample that is removed from its source and tested in an isolated instrument. (See Section 4.)

1.2 This test method is applicable to the measurement of turbidities greater than 1.0 turbidity unit (TU). The upper end of the measurement range was left undefined because different technologies described in this test method can cover very different ranges. The round robin study covered the range of 0-4000 turbidity units because instrument verification in this range can typically be covered by standards that can be consistently reproduced.

1.3 Many of the turbidity units and instrument designs covered in this test method are numerically equivalent in calibration when a common calibration standard is applied across those designs listed in Table 1. Measurement of a common calibration standard of a defined value will also produce equivalent results across these technologies.

Design and
Reporting Unit

Prominent Application

Key Design Features

Typical
Instrument Range

Suggested
Application Ranges

Nephelometric non-ratio (NTU)

White light turbidimeters. Comply with U.S. EPA Method 180.1 for low level turbidity monitoring.

Detector centered at 90° relative to the incident light beam. Uses a white light spectral source.

0.0-40

0.0-40 Regulatory

Ratio White Light turbidimeters (NTRU)

Complies with ISWTR regulations and Standard Method 2130B. Can be used for both low and high level measurement.

Used a white light spectral source. Primary detector centered at 90°. Other detectors located at other angles. An instrument algorithm uses a combination of detector readings to generate the turbidity reading.

0-10 000

0-40 Regulatory
0-10 000 other

Nephelometric, near-IR turbidimeters, non-ratiometric (FNU)

Complies with ISO 7027. The wavelength is less susceptible to color interferences. Applicable for samples with color and good for low level monitoring.

Detector centered at 90° relative to the incident light beam. Uses a near-IR (780-900 nm) monochromatic light source.

0-1000

0-40 Regulatory (non-US)
0-1000 other

Nephelometric near-IR turbidimeters, ratio metric (FNRU)

Complies with ISO 7027. Applicable for samples with high levels of color and for monitoring to high turbidity levels.

Uses a near-IR monochromatic light source (780-900 nm). Primary detector centered at 90°. Other detectors located at other angles. An instrument algorithm uses a combination of detector readings to generate the turbidity reading.

0-10 000

0-40 Regulatory
0-10 000 other

Surface Scatter Turbidimeters (NTU)

Turbidity is determined through light scatter from or near the surface of a sample.

Detector centered at 90° relative to the incident light beam. Uses a white light spectral source.

10-10 000

10-10 000

Formazin Back Scatter (FBU)

Not applicable for regulatory purposes. Best applied to high turbidity samples. Backscatter is common with but not all only probe technology and is best applied in higher turbidity samples.

Uses a near-IR monochromatic light source in the 780-900 nm range. Detector geometry is between 90° and 180° relative to the incident light beam.

100-10 000+

100-10 000

Backscatter Unit (BU)

Not applicable for regulatory purposes. Best applied for samples with high level turbidity.

Uses a white light spectral source (400-680 nm range). Detector geometry is between 90° and 180° relative to the incident light beam.

10-10 000+

100-10 000+

Formazin attenuation unit (FAU)

May be applicable for some regulatory purposes. This is commonly applied with spectrophotometers. Best applied for samples with high level turbidity.

Detector is geometrically centered at 0° relative to incident beam (attenuation). Wavelength is 780-900 nm.

20-1000

More ASTM standard pdf

ASTM D6272-17

ASTM D6272-17

Standard Test Method for Flexural Properties of Unreinforced and Reinforced Plastics and Electrical Insulating Materials by Four-Point Bending

$30.00 $60.00

ASTM F1394-92(2012)

ASTM F1394-92(2012)

Standard Test Method for Determination of Particle Contribution from Gas Distribution System Valves

$43.00 $86.00

ASTM F2137-16

ASTM F2137-16

Standard Practice for Measuring the Dynamic Characteristics of Amusement Rides and Devices

$30.00 $60.00

ASTM C661-15

ASTM C661-15

Standard Test Method for Indentation Hardness of Elastomeric-Type Sealants by Means of a Durometer

$28.00 $56.00