• ASTM F2077-11

ASTM F2077-11

Test Methods For Intervertebral Body Fusion Devices

ASTM International, 07/15/2011

Publisher: ASTM

File Format: PDF

$29.00$58.00


Published:15/07/2011

Pages:9

File Size:1 file , 220 KB

Note:This product is unavailable in Russia, Ukraine, Belarus

Intervertebral body fusion device assemblies are generally simple geometric-shaped devices which are often porous or hollow in nature. Their function is to support the anterior column of the spine to facilitate arthrodesis of the motion segment. This test method outlines materials and methods for the characterization and evaluation of the mechanical performance of different intervertebral body fusion device assemblies so that comparisons can be made between different designs.

This test method is designed to quantify the static and dynamic characteristics of different designs of intervertebral body fusion device assemblies. These tests are conducted in vitro to allow for analysis and comparison of the mechanical performance of intervertebral body fusion device assemblies to specific force modalities.

The forces applied to the intervertebral body fusion assemblies may differ from the complex loading seen in vivo, and therefore, the results from these tests may not directly predict in vivo performance. The results, however, can be used to compare mechanical performance of different intervertebral body fusion device assemblies.

Since the environment may affect the dynamic performance of intervertebral body fusion device assemblies, dynamic testing in a saline environment may be considered. Fatigue tests should first be conducted in air (at ambient temperature) for comparison purposes since the environmental effects could be significant. If a simulated in vivo environment is desired, the investigator should consider testing in a saline environmental bath at 37°C (for example, 0.9-g NaCl per 100-mL water) at a rate of 1 Hz or less. A simulated body fluid, a saline drip or mist, distilled water, or other type of lubrication at 37°C could also be used with adequate justification.

If the devices are known to be temperature and environment dependent, testing should be conducted in physiologic solution as described in 5.4. Devices that require physiologic solution for testing should be tested in the same type solution for comparison purposes.

The location within the simulated vertebral bodies and position of the intervertebral body fusion device assembly with respect to the loading axis will be dependent upon the design, the manufacturer's recommendation, or the surgeon's preferred method for implant placement.

It is well known that the failure of materials is dependent upon stress, test frequency, surface treatments, and environmental factors. Therefore, when determining the effect of changing one of these parameters (for example, frequency, material, or environment), all others must be kept constant to facilitate interpretation of the results.

More ASTM standard pdf

ASTM C216-22

ASTM C216-22

Standard Specification for Facing Brick (Solid Masonry Units Made from Clay or Shale)

$38.00 $76.00

ASTM F1488-09

ASTM F1488-09

Standard Specification for Coextruded Composite Pipe

$34.00 $69.00

ASTM D7591-12(2017)

ASTM D7591-12(2017)

Standard Test Method for Determination of Free and Total Glycerin in Biodiesel Blends by Anion Exchange Chromatography

$24.00 $48.00

ASTM D2584-18

ASTM D2584-18

Standard Test Method for Ignition Loss of Cured Reinforced Resins

$21.00 $42.00