• ASTM F2077-22

ASTM F2077-22

Standard Test Methods for Intervertebral Body Fusion Devices

ASTM International, 07/01/2022

Publisher: ASTM

File Format: PDF

$30.00$60.00


Published:01/07/2022

Pages:11

File Size:1 file , 360 KB

Note:This product is unavailable in Russia, Ukraine, Belarus

1.1''This test method covers the materials and methods for the static and dynamic testing of intervertebral body fusion device assemblies, spinal implants designed to promote arthrodesis at a given spinal motion segment.

1.2''This test method is intended to provide a basis for the mechanical comparison among past, present, and future nonbiologic intervertebral body fusion device assemblies. This test method allows comparison of intervertebral body fusion device assemblies with different intended spinal locations and methods of application to the intradiscal spaces. This test method is intended to enable the user to compare intervertebral body fusion device assemblies mechanically and does not purport to provide performance standards for intervertebral body fusion device assemblies.

1.3''The test method describes static and dynamic tests by specifying force types and specific methods of applying these forces. These tests are designed to allow for the comparative evaluation of intervertebral body fusion device assemblies.

1.4''These tests are designed to characterize the structural integrity of the device and are not intended to test the bone-implant interface.

1.5''This test method does not address expulsion testing of intervertebral body fusion device assemblies (see 1.4).

1.6''Guidelines are established for measuring displacements, determining the yield force or moment, and evaluating the stiffness and strength of the intervertebral body fusion device assemblies.

1.7''Some intervertebral body fusion device assemblies may not be testable in all test configurations.

1.8''The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard, with the exception of angular measurements, which may be reported in terms of either degrees or radians.

1.9''This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.10''This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

More ASTM standard pdf

ASTM D4590-06

ASTM D4590-06

Standard Test Method for Colorimetric Determination of p-tert-Butylcatechol In Styrene Monomer or AMS (alpha-Methylstyrene) by Spectrophotometry

$26.00 $52.00

ASTM D6205-06

ASTM D6205-06

Standard Practice for Calibration of the James Static Coefficient of Friction Machine

$30.00 $60.00

ASTM C1294-07

ASTM C1294-07

Standard Test Method for Compatibility of Insulating Glass Edge Sealants with Liquid-Applied Glazing Materials

$26.00 $52.00

ASTM D4984-06

ASTM D4984-06

Standard Test Method for Carbon Dioxide in Natural Gas Using Length-of-Stain Detector Tubes

$26.00 $52.00