Your shopping cart is empty!
ASTM International, 06/15/2007
Publisher: ASTM
File Format: PDF
$29.00$58.00
Published:15/06/2007
Pages:5
File Size:1 file , 150 KB
Note:This product is unavailable in Russia, Ukraine, Belarus
1.1 This test method covers a procedure for the fatigue testing of metallic femoral hip prostheses used in hip joint replacements. This test method covers the procedures for the performance of fatigue tests on metallic femoral hip stems using a cyclic, constant-amplitude force. It applies to hip prostheses that utilize proximal metaphyseal fixation and are of a modular construct, and it is intended to evaluate the fatigue performance of the modular connections in the metaphyseal filling (that is, proximal body) region of the stem.
1.2 This test method is intended to provide useful, consistent, and reproducible information about the fatigue performance of metallic hip prostheses while held in a proximally fixated manner, with the distal end not held by a potting medium.
1.3 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Standard Specification for Ni-Fe-Cr-Mo-Cu Low-Carbon Alloy (UNS N08904), Ni-Fe-Cr-Mo-Cu-N Low-Carbon Alloys (UNS N08925, UNS N08031, and UNS N08926), and Cr-Ni-Fe-N Low-Carbon Alloy (UNS R20033) Bar and Wire
$30.00 $60.00
Standard Guide for Predicting Neutron Radiation Damage to Reactor Vessel Materials, E 706 (IIF)
$29.00 $58.00
Test Method for Shrinkage Factors of Soils by the Mercury Method
$27.00 $54.00
Standard Practice for Sonic Drilling for Site Characterization and the Installation of Subsurface Monitoring Devices