1.1 This standard specifies procedures for performance verification of multi-axis force platforms commonly used for measuring ground reaction forces during gait, balance and other activities.
1.1.1 This standard provides a method to quantify the relationship between applied input force and force platform output signals across the manufacturer's defined spatial working surface and specified force operating range.
1.1.2 This standard provides definitions of the critical parameters necessary to quantify the behavior of multi-axis force measuring platforms and the methods to measure the parameters.
1.1.3 This standard presents methods for the quantification of spatially distributed errors and absolute measuring performance of the force platform at discrete spatial intervals and discrete force levels on the working surface of the platform.
1.1.4 This standard further defines certain important derived parameters, notably COP (center of pressure) and methods to quantify and report the measuring performance of such derived parameters at spatial intervals and force levels across the working range of the force platform.
1.1.5 This standard defines the requirements for a report suitable to characterize the force platform's performance and provide traceable documentation to be distributed by the manufacturer or calibration facility to the users of such platforms.
1.1.6 Dynamic characteristics and applications where the force platform is incorporated in other equipment, such as instrumented treadmills and stairs, are beyond the scope of this standard.
1.1.7 This standard is written for purposes of multi-axis force platform verification; however the methods and procedures are applicable to calibration of force platforms by manufacturers.
1.2 The values stated in SI units are to be regarded as the standard. Other metric and inch-pound values are regarded as equivalent when required.
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.