ASTM F564-02(2006)e1

Standard Specification and Test Methods for Metallic Bone Staples

ASTM International, 03/01/2006

Publisher: ASTM

File Format: PDF

$32.00$65.00


Published:01/03/2006

Pages:12

File Size:1 file , 220 KB

Note:This product is unavailable in Russia, Ukraine, Belarus

1.1 This specification covers characterization of the design and mechanical function of metallic staples used in the internal fixation of the muscular skeletal system. It is not the intention of this specification to describe or specify specific designs for metallic bone staples.

1.2 This specification includes the following four test methods for measuring mechanical properties of metallic bone staples:

1.2.1 Test Method for Constant Amplitude Bending Fatigue Tests of Metallic Bone Staples-Annex A1.

1.2.2 Test Method for Pull-Out Fixation Strength of Metallic Bone Staples-Annex A2.

1.2.3 Test Method for Soft Tissue Fixation Strength of Metallic Bone Staples-Annex A3.

1.2.4 Test Method for Elastic Static Bending of Metallic Bone Staples-Annex A4.

1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

A1.1.1 This test method covers procedures for the performance of constant amplitude fatigue testing of metallic staples used in internal fixation of the musculoskeletal system. This test method may be used when testing in air at ambient temperature or in an aqueous or physiological solution.

A1.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

A1.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

A2.1.1 This test method covers testing of the hard tissue pull-out fixation strength of metallic staples used in the internal fixation of the musculoskeletal system. This test method may be used with physiologic bone or a synthetic substitute. It may also be used when testing in an aqueous or physiological solution.

A2.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

A2.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

A3.1.1 This test method covers testing the soft tissue fixation strength of metallic staples used in internal fixation of the musculoskeletal system. This test method may be used with physiologic soft tissue and bone or synthetic substitutes for either, or both. This test method may also be used when testing in an aqueous or physiological solution.

A3.1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

A3.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

A4.1.1 This test method covers a procedure for determining the rigidity of bone staples.

A4.1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.

A4.1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

More ASTM standard pdf

ASTM D1047-16

ASTM D1047-16

Standard Specification for Poly(Vinyl Chloride) Jacket for Wire and Cable

$27.00 $54.00

ASTM D2674-72(2021)

ASTM D2674-72(2021)

Standard Methods of Analysis of Sulfochromate Etch Solution Used in Surface Preparation of Aluminum

$23.00 $46.00

ASTM D991-89(2014)

ASTM D991-89(2014)

Standard Test Method for Rubber Property-Volume Resistivity Of Electrically Conductive and Antistatic Products

$30.00 $60.00

ASTM E45-18a

ASTM E45-18a

Standard Test Methods for Determining the Inclusion Content of Steel

$29.00 $58.00