• UL 61010-2-091

UL 61010-2-091

Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-091: Particular requirements for cabinet X-Ray systems

Underwriters Laboratories, 08/08/2014

Publisher: UL

File Format: PDF

$125.00$251.64


Published:08/08/2014

Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-091: Particular Requirements for Cabinet X-Ray Systems
UL 61010-2-091 (IEC 61010-2-091:2012)
1 Scope and object
This clause of Part 1 is applicable, except as follows:
1.1.1 Equipment included in scope
Replacement:Replace the text with the following:A CABINET X-RAY SYSTEM is a system that contains an X-ray tube installed in a cabinet which, independently of existing architectural structures except the floor on which it may be placed, is intended to contain at least that portion of a material being irradiated, provide radiation attenuation, and exclude personnel from the interior during generation of X-radiation.These CABINET X-RAY SYSTEMS are used in industrial, commercial, and public environments, for example, to inspect materials, to analyze materials, and to screen baggage.

1.1.2 Equipment excluded from scope
Addition:Add the following new items to the list:
aa) equipment intended to apply X-radiation to humans or animals;
bb) equipment incorporating an X-ray tube but not incorporating complete shielding against X-radiation hazards, such as:
  • equipment intended to be used within a shielded room which excludes personnel during operation;
  • equipment intended to be used with separate portable or temporary shielding;
  • equipment intended to produce an emerging beam of X-radiation.
1.2.1 Aspects included in scope
Addition:Add the following text at the end of the first paragraph:This part of IEC 61010 specifies requirements for the design and methods of construction of CABINET X-RAY SYSTEMS to provide adequate protection for OPERATORS, bystanders, trained service personnel, and the surrounding area against unintentionally-emitted X-radiation and from mechanical HAZARDS related to their conveyors.

More UL standard pdf

UL 1453 CRD

UL 1453 CRD

Standard for Electric Booster and Commercial Storage Tank Water Heaters (Certification Requirement Decision, 10/3/2016)

$112.00 $224.84

UL Subject 1573

UL Subject 1573

Standard for Stage and Studio Luminaires and Connector Strips (CSDS Proposal, 10/3/2016)

$139.00 $278.34

UL Subject 979

UL Subject 979

Standard for Water Treatment Appliances (CSDS Proposal, 10/3/2016)

$114.00 $228.75

UL 711 CRD

UL 711 CRD

Rating and Fire Testing of Fire Extinguishers (Certification Requirement Decision, 10/3/2016)

$129.00 $258.92